Section 3 • Programming Instructions
36
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
(C) Repeat steps 5(A) and 5(B) to
set the digits in the tens and ones
sections.
Volume to be infused is
established.
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS button
is pressed, the display moves
backward one screen to
enable review or correction.
NOTE
: Bag volume limits are 1 to 1,000 ml.
Pump returns to pause mode.
One beep
followed by
two beeps
Pause icon flashes.
NOTE
: To confirm program settings, access “RX” by pressing and releasing the BOLUS button.
After scrolling through the programmed settings, the pump will automatically return to
pause mode.
3.2.2.3
Moving Backward in Display While Programming
The user can move backward in the display. This functions as follows:
•
Pushing and holding the BOLUS button held until a beep is heard causes the currently
blinking value to be accepted (i.e., stored in memory) and the value to the left on the
display to start to blink so that it can be changed.
•
If the leftmost value is blinking and the BOLUS button is pushed and held until a beep
is heard, then the value is accepted (i.e., stored in memory) and the display changes
the preceding parameter. For example if the left most value of the bolus lockout time
(hours) is blinking and the BOLUS button is pushed and held held until a beep is heard,
the hours value is accepted and the display changes to the previous parameter (bolus
volume) and the right most digit blinks (tenths digit).
•
If the basal rate left most digit (tens digit: 0,1X, 2X) is blinking and the BOLUS button
is held until a beep is heard, then the value is accepted and the pump shows the mode
selected “PCA”.
•
If the mode (PCA) is blinking and the BOLUS button is held until a beep is heard, the
mode is accepted and the pump displays the first parameter for that mode and the left
most value is blinking.
NOTE:
If the mode is blinking, then the BOLUS button functions the same as the RUN/
PAUSE button.