
Section 1 • Introduction
25
To power off the pump:
Step #1
If the pump is in run mode place the pump in pause mode by pushing
the RUN/PAUSE button (see Section 4.2).
Step #2
Rotate the battery cover counter-clockwise until the (
O
) mark on the
cover lines up with the
( I )
mark on the pump (see illustration at left).
2.8
Pump Power On and Off
CAUTION:
Always place the pump in pause mode prior to turning the pump off. Failure to
do so may cause the therapy to be delayed and/or history to be lost. The pump
will not sound an alarm if the pump is not placed in pause mode prior to being
turned off.
To power on the pump:
Rotate the battery cover clockwise until the
( I )
mark on the cover lines up with the
( I )
mark
on the pump.
NOTE:
After the power-on self-test, the pump will beep twice and go into pause mode.
The clinician can then program the pump or resume current infusion settings.
Power OFF
counter-
clockwise
Power ON