Section 3 • Programming Instructions
31
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS
button is pressed, the
display moves backward
one screen to enable review
or correction.
NOTE
: If bolus volume is set to “0,” this step will be skipped.
NOTE
: The limits for lockout time are 00:01 to 24:00 hh:mm.
6
Set volume to be infused:
(A) Press and release the RUN/
PAUSE button to increase the
number or press and release the
BOLUS button to decrease the
number.
“ml” is displayed, and the
bag icon and first section
flash.
(B) When the desired number
appears in the display, press and
hold the RUN/PAUSE button until
one beep is heard to accept the
setting.
One beep
NOTE: Volume to be
infused is displayed in ml.
(C) Repeat the steps 6(A) and 6(B)
to set the digits in the tens and ones
section.
“ml” is displayed. The bag
icon and respective digit
flash.
Volume to be infused is
established.
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS
button is pressed, the
display moves backward
one screen to enable review
or correction.
NOTE
: Bag volume limits are
1 to 1000 ml.
Pump returns to pause mode.
One beep
followed by
two beeps
Pause icon flashes.
Note: To confirm program
settings, access “RX” by
pressing and releasing
the BOLUS button. After
scrolling through the
programmed settings, the
pump will automatically
return to pause mode.
ml
ml