Section 3 • Programming Instructions
27
3.2
Program Options
To enter program mode:
Step #1
Place the pump in pause mode. If the pump is not on, turn it on (see Section 2.7).
Step #2
Press and hold both the RUN/PAUSE button and the BOLUS button for two seconds until one
beep is heard.
“PRO” will momentarily appear in the pump display, after which the pump mode -- PIb, PCA,
or P+P -- will flash in the pump display, indicating that the pump mode can be changed. The
remaining steps for programming the pump are described in Sections 3.2.1 through 3.2.3.
Changes to previous programming screens may be made during programming by pressing the
BOLUS button for two seconds until a beep is heard. The immediately previous screen will be
displayed. This may be repeated until the initial screen is reached.
The pump can be reprogrammed during infusion, without clearing the infusion history, by
placing the pump in pause mode and then entering program mode. See Section 3.2.1 through
3.2.3 for programming steps.
The pump can be reprogrammed after clearing the infusion history (see Section 5.2 – “Clearing
Pump Infusion History”) by entering program mode and selecting the desired parameters.
When attempting to program/reprogram the pump, “LOC” will appear in the pump display if
the pump program is in lockout mode (see Section 6 – “Patient Lockout”).
The pump program must be unlocked for programming/reprogramming and clearing the
infusion history.
WARNING:
The pump should be programmed by a healthcare provider who has been
trained to program and set up the pump. Patients should be instructed not to
program the pump or attempt to change the program.
CAUTION:
After clearing history or reprogramming the pump a bolus (in PCA mode) or
dose (in PIb or P+P modes)
is
available immediately after starting the infusion.
NOTE:
The remote BOLUS switch cannot be used to program the pump.