Section 3 • Programming Instructions
41
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS button
is pressed, the display moves
backward one screen to enable
review or correction.
NOTE
: The maximum limit for the lockout time is dependent upon the dose volume, dose interval
time and/or the bolus volume programmed.
8
Set volume to be infused:
(A) Press and release the RUN/
PAUSE button to increase the
number or press and release the
BOLUS button to decrease the
number.
“ml” is displayed, and bag icon
and first digit to the left flash.
(B) When the desired number
appears in the display, press and
hold the RUN/PAUSE button or
the BOLUS button until one beep is
heard to accept the setting.
One beep
NOTE: Volume to be infused is
displayed in ml.
(C) Repeat steps 7(A) and 7(B) to
set the digits in the tens and ones
sections.
“ml” is displayed. Bag icon and
respective digit flashes.
Volume to be infused is
established.
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS button
is pressed, the display moves
backward one screen to enable
review or correction.
NOTE
: Bag volume limits are 1 to 1,000 ml.
Pump returns to pause mode.
One beep
followed by
two beeps
Pause icon flash
es.
NOTE
: To confirm program settings, access “RX” by pressing and releasing the BOLUS button. After
scrolling through the programmed settings, the pump will automatically return to pause
mode.
3.2.3.3
Moving Backward in Display While Programming
The user can move backward in the display. This functions as follows:
•
Pushing and holding the BOLUS button held until a beep is heard causes the currently
blinking value to be accepted (i.e., stored in memory) and the value to the left on the
display to start to blink so that it can be changed.
ml
ml