Section 4 • Operating Instructions
44
If the BOLUS button is pushed within 60 seconds of placing the pump in run mode after
changing the settings or clearing the history, the dose is delivered immediately instead of
waiting for the dose interval to elapse.
4.1.3
P+P Mode Start Infusion
The first dose is delivered after the dose interval and the first bolus is not available until after a
lockout time has expired (see Section 1.4.1 for flow patterns), with the following exception:
If the BOLUS button is pushed within 60 seconds of placing the pump in run mode after
changing the settings or clearing the history, the dose is delivered immediately instead of
waiting the length of the dose interval, if a dose has been programmed. If there is no dose
programmed, the first bolus can be requested after the lockout time has elapsed
.
4.2
Pause Infusion
To pause the infusion, press and release the RUN/PAUSE button. The
pump will beep two times, the green run light will stop blinking, and
the pause mode icon (
I I
) will flash in the pump display. If left in pause
mode, the pump will beep two times every four minutes.
CAUTION:
Always place the pump in pause mode prior to turning
the pump off. Failure to do so may cause the therapy
to be delayed and/or history to be lost. The pump will
not sound an alarm if the pump is not placed in pause
mode prior to being turned off.
NOTE:
Pausing the pump temporarily stops the infusion. While
in pause mode, the infusion is delayed. This allows for
changing the cassette, solution container, or batteries.
4.3
Resume Infusion
To resume the infusion from pause mode, press and release the RUN/
PAUSE button.
The green run light (inside the BOLUS button) will start to blink, the
“ml” icon and the volume infused will appear in the pump display.
The infusion will resume at the same point at which the pump was last
placed in pause mode.
4.4
Silence Alarm
To silence an alarm, press and release the RUN/PAUSE button. When
the alarm has been silenced, the pump will remain in pause mode.
Once the cause of the alarm has been corrected, resume the infusion
(see Section 4.3).