Section 1 • Introduction
24
Step #2
Remove the cover and insert the batteries according to the
illustrations at left.
Step #3
Place the battery cover onto the pump as illustrated to the left
(OFF symbol (
O
) on the cover slightly to the left of the
( I )
mark
on the pump).
Step #4
Rotate the battery cover clockwise to the OFF (
O
) position.
Insert batteries and replace cover
2.7.2
Battery Replacement
WARNING:
When reinstalling batteries, always verify correct program settings before
restarting infusion.
To replace batteries:
CAUTION:
Verify that the pump is in pause mode before removing the batteries. Failure to
do so may cause loss of timing and a delay in therapy. The pump will not sound
an alarm if it is turned off without being placed in pause mode.
Step #1
If the pump is in run mode place the pump in pause mode by
pushing the RUN/PAUSE button (see Section 4.2).
Step #2
Rotate the battery cover counter-clockwise slightly to the left of
the OFF (
O
) position, until the cover stops or meets resistance.
Step #3
Remove the cover and insert the batteries according to the
instructions in Section 2.6.1.
Step #4
Power on the pump according to the instructions in Section 2.6.3.
After the batteries are replaced and the pump is powered on, the pump will return to pause
mode.
Press and release the RUN/PAUSE button to review the current program settings.
Press and release the RUN/PAUSE button a second time to resume the current infusion
program.