Alcatel Vacuum Technology France - User’s Manual ATH 1300 M/MT - ATH 1600 M/MT
C 60
1/1
GB 00750 - Edition 04 - March 00
C 60
«External safety» contact operation
If the «EXT. SAFETY» contact on the «Inputs/Outputs»
connector
(see B 91)
is open:
- the controller display
indicates:
- the pump is stopped and the air inlet valve opens if
the air inlet valve parameter is setted.
To restart the pump, closed the «EXT. SAFETY» contact
and start the pump:
- in local mode: push the START key;
- in remote control mode: open «START» contact then,
close it
(see B 91)
.
D 0 6
: E X T . S A F E T Y
This operation mode is always operational even
on “Local“ or “Remote“ mode.
CAUTION!