GB 00754 - Edition 08 - Januar
y 04
D 20
Alcatel Vacuum Technology France - User’s Manual ATH 1300 M/MT - ATH 1600/M/MT
Diagnosis and troubleshooting
6/8
Default is not indicated
by the controller
■
Check that the controller is
powered.
■
Check that the power voltage
is between 100 V -15% and
120 V +10% or between
200 V -15% and 240 V +10%.
■
Check the fuses: 2 fuses
(16 A) under the controller.
■
Contact the Customer Service.
■
Pump stopped, reinitialize
the controller with the main
switch (0/1).
■
Contact the Customer Service.
■
Change the cable link
between the pump and the
controller.
■
Contact the Customer Service.
■
Check that the pump is rigidly
fixed to the frame;
■
Check that there are no
vibrations on the pump frame;
■
Check that the attachment
are correctly fastened
(Problem can be caused by an
anti-vibration flagstone);
■
Contact the Customer Service.
I
NCIDENT
C
AUSE
R
EMEDY
No event occurs
after power on
Isolate the product from
the various energy
sources before any
maintenance operations.
No mains current
The ATH only starts-up in
STANDBY mode
Electronic problem
The motor power load
stays below 100W
Electronic problem
Up to 7000 rpm,
the pump starts to vibrate
Mechanical problem
Содержание Adixen ATH 1300 M
Страница 1: ...ATH 1300 1600 M MT MAGLEV HYBRID TURBOMOLECULAR PUMPS User s Manual...
Страница 24: ......
Страница 26: ......
Страница 32: ......
Страница 44: ......
Страница 50: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......
Страница 70: ......
Страница 82: ......
Страница 88: ......
Страница 96: ......
Страница 98: ......
Страница 112: ......
Страница 114: ......
Страница 116: ......
Страница 121: ...G 100 1 1 GB 02326 Edition 01 January 04 Alcatel Vacuum Technology France User s Manual ATH 1300 M MT ATH 1600 M MT...
Страница 122: ......