B 100
Alcatel Vacuum Technology France - User’s Manual ATH 1300 M/MT - 1600 M/MT
1/1
GB 00741 - Edition 05 - Januar
y 03
"Rem." connector wiring
“Rem.“ connector at
the rear of the controller
Used to monitor the selected parameter
(see ANALOG OUT menus C 30).
1
2
3
4
5
9
10
11
12
6
7
8
14
13
15
RS485 V(+)
RS485 V(-)
0V
Analog output (0-10V)
Analog.
Output
8 - 15
Analog output signal
The signal is transmitted between terminal 8 and 15 of
the remote connector.
Four values can be used to plot curves:
Pump rotation speed* 360 rpm (OV) to 60000 rpm
(10V)
Pump temperature
0°C
(OV) to
100°C
(10V)
Controller temperature
0°C
(OV) to
100°C
(10V)
Current
0mA
(OV) to
8700mA
(5.8V)
Factory configuration is setted on
speed.
0
10
Voltage (V)
0
0
Controller temperature (
°
C)
Speed (rpm)
Current (mA)
0
Pump temperature (
°
C)
0
100
100
60000 rpm
15000
*ATH 1600 39000 rpm (6.5 V)
*ATH 1300
34000 rpm (5.66 V)
*ATH 2300
31000 rpm (5.16 V)
Содержание Adixen ATH 1300 M
Страница 1: ...ATH 1300 1600 M MT MAGLEV HYBRID TURBOMOLECULAR PUMPS User s Manual...
Страница 24: ......
Страница 26: ......
Страница 32: ......
Страница 44: ......
Страница 50: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......
Страница 70: ......
Страница 82: ......
Страница 88: ......
Страница 96: ......
Страница 98: ......
Страница 112: ......
Страница 114: ......
Страница 116: ......
Страница 121: ...G 100 1 1 GB 02326 Edition 01 January 04 Alcatel Vacuum Technology France User s Manual ATH 1300 M MT ATH 1600 M MT...
Страница 122: ......