B 120
Detailed description
of RS232 and RS485 commands
3/8
GB 00743 - Edition 04 - June 98
Alcatel Vacuum Technology France - User’s Manual ATH 1300 M/MT - ATH 1600 M/MT
NSP
Syntax
Result
OPT
Syntax
Result
Switches the speed set point to the nominal speed value
#adr
NSP
<CR>[<LF>]
#adr,OK
The speed set point for the pump is set to its nominal value.
Used to select possible user choices
#adr
OPT
01,n<CR>[<LF>]
choice of parameters on the analog output:
n = 0 : real pump speed
n = 1 : pump current
n = 2 : temperature of pump body
n = 3 : temperature of internal electronics
#adr
OPT
02,n<CR>[<LF>]
choice of temperature unit
n = 0 : degrees Centigrade
n = 1 : degrees Fahrenheit
#adr
OPT
11,n<CR>[<LF>]
buzzer
n = 0 : without
n = 1 : with
#adr
OPT
14,n<CR>[<LF>]
remote mode
n = 0 : off
n = 1 : on
#adr,OK
The choice of the temperature unit affects the results of
the DLR and STA strings and the display (if cabinet fitted).
Содержание Adixen ATH 1300 M
Страница 1: ...ATH 1300 1600 M MT MAGLEV HYBRID TURBOMOLECULAR PUMPS User s Manual...
Страница 24: ......
Страница 26: ......
Страница 32: ......
Страница 44: ......
Страница 50: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......
Страница 70: ......
Страница 82: ......
Страница 88: ......
Страница 96: ......
Страница 98: ......
Страница 112: ......
Страница 114: ......
Страница 116: ......
Страница 121: ...G 100 1 1 GB 02326 Edition 01 January 04 Alcatel Vacuum Technology France User s Manual ATH 1300 M MT ATH 1600 M MT...
Страница 122: ......