B 40
Alcatel Vacuum Technology France - User’s Manual ATH 1300 M/MT - ATH 1600 M/MT
Air inlet valve connection
(option)
1/2
GB 00734 - Edition 07 - March 03
Function
Vacuum connection
Without purge device
With purge device
The air inlet valve is calibrated to reset the volume of
the pump to atmospheric pressure.
When the pump is isolated (at inlet and exhaust) the rotor
slow down efficiency is increased.
If the venting time is setted, the reset to atmospheric
pressure takes place when the pump is stopped or when
faults are registered on the controller
(see C 50)
.
Air inlet valve is delivered with a DN 16 long nipple not
used in this installation.
Install the valve on the DN 16 fitting of the pump.
The valve must be connected to an inert gas line which
can be for example dry nitrogen (Pressure between 1 and
1.5 bars absolute)
(see B 50 for nitrogen characteristics)
.
Connect the inert gas line with a rigid stainless stell or
flexible pipe (ext. diam 1/4“).
See B 50
.
Blank-off frange
Inert gas
Содержание Adixen ATH 1300 M
Страница 1: ...ATH 1300 1600 M MT MAGLEV HYBRID TURBOMOLECULAR PUMPS User s Manual...
Страница 24: ......
Страница 26: ......
Страница 32: ......
Страница 44: ......
Страница 50: ......
Страница 62: ......
Страница 64: ......
Страница 68: ......
Страница 70: ......
Страница 82: ......
Страница 88: ......
Страница 96: ......
Страница 98: ......
Страница 112: ......
Страница 114: ......
Страница 116: ......
Страница 121: ...G 100 1 1 GB 02326 Edition 01 January 04 Alcatel Vacuum Technology France User s Manual ATH 1300 M MT ATH 1600 M MT...
Страница 122: ......