IECEx (Zone 2 and 22 purging system
installation)
230
GasEye Cross Duct SG and MG
Operating instructions, 05/2021
IECEx Marking
Ex op is pzc IIC T6 Gc
Ex op is pzc IIIB T85°C Dc
IECEx KDB 19.0004X
Special conditions for safe use
External parts made of plastic should be cleaned with a damp cloth, with the
addition of antistatic fluids.
Enclosure should be installed in a way that prevents electrostatic charging, in
accordance with the instructions.
Maximum inlet pressure to the containment system should not exceed 2 bar.
System power must not be restored after the enclosure has been opened until
combustible gas/dust accumulations within the enclosure have been removed
Figure 220. Gas analyzer identification plate – IECEx 2/22.