WebServer application
GasEye Cross Duct
Operating instructions, 05/2021
171
Figure 170. Dout parameters window.
8.4.1.12. IO -> Aout (Analog output configuration)
Analog value is scaled to a defined range and set to the specific output.
Formula used to calculate output value:
𝐴𝑜𝑢𝑡 = 4 +
𝑀𝐴𝑁𝑈𝐴𝐿_𝑉𝐴𝐿𝑈𝐸 − 𝑆𝐶𝐴𝐿𝐸_𝑀𝐼𝑁
𝑆𝐶𝐴𝐿𝐸_𝑀𝐴𝑋 − 𝑆𝐶𝐴𝐿𝐸_𝑀𝐼𝑁
∗ 16
Analog outputs parameters
Name
Description
AOUT.FORCE_MANUAL_
MODE_ENABLE
Switch all analog output channels to
manual
mode
ignoring
SELECT_SIGNAL value
AOUT.SCALE_ENABLE
Enable/disable calculation of scaling
value
from
range
SCALE_MIN,
SCALE_MAX (formula above)
AOUT.CALIBRATED
Enable/disable calibration factors
Содержание GasEye