10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in
Table 9 on
page 32
. This can be done using Limit Lines, select
Marker/Analysis
>
Analysis
>
Limit
Test
>
Limit Test ON
and
Limit Line ON
>
Global pass/fail display ON
, select
Show
Table.
The trace should be above the limit. PASS will be displayed on the screen if the
limit lines are used.
11.Repeat
step 9
and
step 10
for Cal Sets “999.1” thru “999.12” (12-Port).
NOTE
For Ports 3-12 the response from 10 MHz to 500 MHz is normal due to the
PNA-X Couplers in comparison to the U3022AE10 bridges. The bridges have
more gain in the coupled RF path.