User’s Guide U3025-90001
17
U3025AE10
Controlling the System with E8363/4C and N5230C
Controlling the System with E8363/4C and N5230C
This section will describe how to setup and operate the U3025AE10 Multiport Test Set
with the E8363/4C and N5230C 2-Port PNA Network Analyzer.
NOTE
The internal firmware of the Agilent E8363/4C and N5230C has not been
modified for this Test Set option. Power levels may differ from those indicated
on the PNA when the Test Set is connected.
The U3025AE10 Multiport Test Set is considered a “slave” instrument. A E8363/4C or
N5230C must be used to control the Test Set.
• PNA Multiport Mode Option 551 requires firmware revision
≥
A.08.04.10. From the
PNA drop-down menu select
Help
>
About Network Analyzer
to see the options
installed and the installed firmware revision, refer to
Figure 12
.
Figure 12
N5230C and E8363/4C Options
The following Test Set file must be installed into network analyzer file directory:
c:\Program Files\Agilent\Network Analyzer\testsets
2-Port PNA-X requires Test Set: u302xae10_pnax_p2.tsx
2-Port PNA or PNA-L requires Test Set: u302xae10_p2.tsx
Visit our website for firmware revision and
downloads.http://na.tm.agilent.com/pna/firmware/firmware.htm
This Manual:http://www.manuallib.com/agilent/u3025ae10-test-set-user-guide.html