![Agilent Technologies U3025AE10 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/agilent-technologies/u3025ae10/u3025ae10_user-manual_2868216016.webp)
User’s Guide U3025-90001
9
U3025AE10
General Specifications
Typical Reflection Tracking
Specifications for the Test Set are typical. System performance for the PNA and Test Set
are only characteristic and are intended as non-warranted information.
NOTE
Typical specifications are based on 1 to 2 units performance, refer to
Table 4
.
Table 4
Typical Reflection Tracking Loss
1,
2
1.
Reflection tracking loss is generally source path loss + receiver path loss.
2.
Does not include margin of ± 2 dB, or PNA accuracy ± 1.5 dB up to 40 GHz
(± 2 dB 40 to 50 GHz).
Frequency
Standard (dB)
(typical)
Option 001 (dB)
(typical)
10 MHz to 45 MHz
+3.0
+2.5
45 MHz to 500 MHz
+6.5
+6.0
500 MHz to 2 GHz
+4.0
+3.5
2 GHz to 10 GHz
+2.0
+1.5
10 GHz to 20 GHz
0
−
0.5
20 GHz to 30 GHz
−
3.0
−
8.0
30 GHz to 40 GHz
−
4.5
−
9.0
40 GHz to 45 GHz
−
7.0
−
10.0
45 GHz to 50 GHz
−
8.5
−
12.0
This Manual:http://www.manuallib.com/agilent/u3025ae10-test-set-user-guide.html