User’s Guide U3025-90001
33
U3025AE10
Operational Check Procedures
Equipment Required
The Agilent U3025AE10 requires that the user be familiar with the equipment and
components listed in
Table 10
.
This section provides an equipment list and setup of the PNA and Test Set.
E8363/4C and N5230C Procedure
1. Connect the Test Set to the 2-Port E8363/4C or N5230C 2-Port PNA as shown in
Figure
10 on page 15
, and
Table 5 on page 15
.
2. Turn On the Test Set.
3. On the E8363/4C and N5230C, select
Response
>
CAL
>
Manage Cals
>
Cal Set
. Delete
or Rename any Cal Sets titled “999.1” thru “999.12” (12-Port), although it is unlikely
that you will find Cal Sets with these names.
4. Verify that the PNA is in 12-Port. See the bottom of the measurement window.
a. If only four S-Parameters are listed, select
Utility
>
System
>
Configure
>
Multiport
Capability
. Select
Restart as multiport PNA with this test set
and select
U302xAE10
(12-Port).
5. Press
Preset.
6. Verify that the
[Stop Frequency]
is set to
[50 GHz]
.
7. Verify that the
[Start Frequency]
is set to
[10 MHz]
.
8. Verify that the
[Power]
is to set to
[
−
17 dBm]
. If not, press
[Power]
>
Power Level
and
enter
[
−
17 dBm].
9. Select
Response
>
Avg > IF Bandwidth
>
100 Hz
.
10.Select
Stimulus
>
Sweep
>
Number of Points
>
401
.
11.Connect the ECal module to the network analyzer USB port on the front or rear panel.
This procedure assumes you are using a ECal. If you are not, see
“1-Port Calibration
Procedure”
step 2
on
page 34
.
12.Allow the ECal module, Test Set and PNA to warm up for a minimum of 30 minutes.
Table 10
Equipment List
Description
Qty
E8364C 2-Port Network Analyzer (Option 014 or 551)
or
N5230C 2-Port Network Analyzer (Option 525 or 551)
1
N4693A ECal Module with a female connector, or a 2.4 mm Cal kit
(85056B, 85057D, etc.)
1
Set of interconnect cables (PNA and Test Set), see
“RF Interface Cable
Connections” on page 15
.
1
This Manual:http://www.manuallib.com/agilent/u3025ae10-test-set-user-guide.html