Operational Check Procedures
Operational Check Procedures
This section provides the test calibration procedure to confirm the U3025AE10 and PNA
operational performance. The operation verification limits provided ensure that your
U3025AE10 and PNA are operating properly.
Verification Limits
Specifications for the U3025AE10 Multiport Test Set are typical. System performance for
the PNA and Test Set are only characteristic and intended as non warranted information.
A functional certificate is provided for the U3025AE10 only.
It is recommended that you return your instrument to Agilent Technologies for servicing or
repair if the Test Set and PNA performance exceed the operational verification limits.
NOTE
Typical specifications are based on 1 to 2 units performance.
NOTE
If you suspect that your 12-Port configuration is not operating properly,
ensure that all front RF jumper interconnect cables are correctly attached.
Table 9
Operational Check Reflection Tracking Loss
Frequency
Standard (dB)
Option 001 (dB)
10 MHz to 45 MHz
−
0.5
−
1.0
45 MHz to 500 MHz
+3.0
+2.5
500 MHz to 2 GHz
+0.5
0
2 GHz to 10 GHz
−
1.5
−
2.0
10 GHz to 20 GHz
−
3.5
−
4.0
20 GHz to 30 GHz
−
6.5
−
11.5
30 GHz to 40 GHz
−
8.0
−
12.5
40 GHz to 45 GHz
−
10.5
−
14.0
45 GHz to 50 GHz
−
12.5
−
16.0