![Agilent Technologies U3025AE10 Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/agilent-technologies/u3025ae10/u3025ae10_user-manual_2868216022.webp)
User’s Guide U3025-90001
15
U3025AE10
System Setup E8363C, E8364C or N5230C
RF Interface Cable Connections
Figure 10
illustrates the setup configuration of the U3025AE10 Multiport Test Set and
how it should be configured to the E8364C 2-Port and N5230C PNA Network Analyzer.
1. Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x8)
on the PNA. The RCVR R1 - R2 to SOURCE OUT reference loop jumpers (x2) remain on
the front panel.
2. Connect the RF cables supplied with this option between the network analyzer and the
U3025AE10 Test Set. Torque each cable to 8 in-lb. Refer to
Table 5
and
Figure 10 on
page 15
.
CAUTION
Each end of the interconnect RF cables have a different length from the bend.
When connecting the RF Interconnect cables be sure that the longer end from
the bend is connected to the PNA.
CAUTION
Over torque will cause damage to the Test Set and may cause connectors to
spin or become loose.
Figure 10
12-Port Setup Configuration
Table 5
12-Port Cable Connection
RF Cables
From (PNA)
To (Test Set)
U3025-20044
SOURCE OUT
SOURCE OUT
U3025-20044
CPLR THRU
CPLR THRU
U3025-20045
CPLR ARM
CPLR ARM
U3025-20045
RCVR A IN, B IN
A IN, B IN
44 (x4)
45
44
45 (x4)
This Manual:http://www.manuallib.com/agilent/u3025ae10-test-set-user-guide.html