82
U2701A/U2702A Modular Oscilloscopes User’s Guide
3
Characteristics and Specifications
Vertical system: oscilloscope channels (continued)
Dynamic range
±
4 div
Input impedance
1 M
Ω
:
≈
16 pF
Coupling
AC, DC, Ground
BW limit
≈
25 MHz
Standard probes
10:1 Passive Probe 150 MHz 1.2 m
10:1 Passive Probe 300 MHz 1.2 m
ESD tolerance
±
2 kV
Noise peak-to-peak
3 mVpp
DC vertical offset accuracy
≤
200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value;
> 200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value
DC vertical gain accuracy
±4.0% of Full Scale
Single cursor accuracy
±{DC vertical gain ac DC vertical offset ac 0.2% full scale (~1/2 LSB)}
Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{4.0% (80 mV) + 0.1(10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2% (80 mV)}
= ±6.385 mV
Dual cursor accuracy
±{DC vertical gain ac 0.4% full scale (~1 LSB)}
Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{4.0% (80 mV) + 0.4% (80 mV)} = ±3.52 mV
Horizontal
Range
1 ns/div to 50 s/div
Time base accuracy
20 ppm
Delay range
Pre-trigger: –100%
Post-trigger: +100%
Modes
Main, roll, XY
XY
Yes
Reference position
Center
Modes
Normal, single, auto trigger
Holdoff time
60 ns
U2700A_Modular_Scope.book Page 82 Tuesday, May 6, 2008 9:32 AM