![Agilent Technologies U2701A Скачать руководство пользователя страница 97](http://html1.mh-extra.com/html/agilent-technologies/u2701a/u2701a_user-manual_2868720097.webp)
Characteristics and Specifications
3
U2701A/U2702A Modular Oscilloscopes User’s Guide
81
Characteristics
All characteristics are typical performance values and are not
warranted. Characteristics are valid after a 30- minute warm- up
period and within ±10°C from firmware calibration temperature.
Acquisition: oscilloscope channels
Real time sample rate
2 channels interleaved
Each channel
1 GSa/s
500 MSa/s
Standard memory depth
2 channels interleaved
Each channel
Normal
32 Mpts
16 Mpts
Single-shot
64 Mpts
32 Mpts
Vertical resolution
8 bits
Peak detection
Yes
Averaging
Any number from 1 to 999
Filter
Sin(x)/x interpolation for time base 1 ns to 100 ns
Sweep modes
Auto, normal, single
Vertical system: oscilloscope channels
Scope channels
U2701A/U2702A: Ch 1 and 2 simultaneous acquisition
AC coupled
U2701A: 3.5 Hz to 100 MHz
U2702A: 3.5 Hz to 200 MHz
Calculated rise time
(= 0.35/bandwidth)
U2701A: 3.5 ns
U2702A: 1.75 ns
Single-shot bandwidth
U2701A: 100 MHz
U2702A: 200 MHz
Range
2 mV/div to 5 V/div (1 M
Ω
)
Maximum input
CAT I 30 Vrms, 42 Vpk
Offset range
±
4 div
U2700A_Modular_Scope.book Page 81 Tuesday, May 6, 2008 9:32 AM