![Agilent Technologies U2701A Скачать руководство пользователя страница 87](http://html1.mh-extra.com/html/agilent-technologies/u2701a/u2701a_user-manual_2868720087.webp)
Scope Features and Functions
2
U2701A/U2702A USB Modular Oscilloscopes User’s Guide
71
Acquisition Mode
The figure below shows the Acquisition Mode on the soft front
panel.
The U2701A/U2702A scopes have the following acquisition
modes.
To reduce the displayed random noise, select
Average
acquisition. This mode will decrease the screen refresh rate.
To avoid waveform aliasing, select
Peak Detect
acquisition.
Peak Detect mode captures the maximum and minimum values
of a waveform over multiple acquisitions.
Table 2-9
List of Acquisition Modes
Menu
Settings
Comments
Mode
Normal
Normal acquisition mode is for most waveforms
with normal decimating at slower sweep speeds,
no averaging.
Peak Detect
Peak Detect acquisition mode is
for displaying
narrow pulses that occur
infrequently (at
slower sweep speeds).
Average
Average acquisition mode is for reducing noise
and increasing resolution (at all sweep speeds
without bandwidth or rise time degradation).
U2700A_Modular_Scope.book Page 71 Tuesday, May 6, 2008 9:32 AM