Getting Started
1
U2701A/U2702A USB Modular Oscilloscopes User’s Guide
3
Warm-up Procedures
Before performing any measurements, ensure that oscilloscope
has gone through the following warm- up procedures,
1
Connect the oscilloscope to the PC using a USB cable.
2
Power- up the oscilloscope.
3
Run the Agilent Measurement Manager.
4
Toggle channel 1 or channel 2 on.
5
Click on the Run button.
Test Considerations
For optimum performance, all procedures should comply with
the following recommendations,
•
Ensure that the ambient temperature is stable.
•
Ensure that the ambient relative humidity is less than 80%.
•
Allow a 30 minutes warm- up period before calibration.
•
Keep all the test connection cables as short as possible.
N O T E
Under standalone use, users are only allowed to measure up to
CAT1 30 Vrms maximum.
For high voltage measurement up to CAT1 300 Vrms, users must
install the L-mount kit on the U2701A/U2702A. Before plugging
into the instrument chassis, ensure that the L-Mount kit installed
on your modular oscilloscope is screwed to the instrument
chassis for chassis grounding purpose (see
). It is required to use the provided 10:1 probes
(N2862A/N2863A) for high voltage measurements.