Scope Features and Functions
2
U2701A/U2702A USB Modular Oscilloscopes User’s Guide
51
X and Y Cursors
To enable Cursors control, the Manual option on the Markers
property must be selected as shown below.
• X cursors
places two cursors on the X- Axis of the waveforms
to measure the time difference between the two cursors (X2
minus X1). Delta X denotes the time difference.
• Y cursors
places two cursors on the Y- Axis of the waveforms
to measure the voltage difference between the two cursors
(Y2 minus Y1). Delta Y denotes the voltage difference.
To configure Auto marker
The Auto marker function will automatically place indicators on
the waveforms displayed on the graph to show the selected
measurements.
1
Obtain and establish a stable signal on the waveform
acquisition graph.
2
Click on the Measurements & Cursors button and select Auto
at the Markers property.
3
Proceed to select your desired measurements on the
Measurements panel.
4
Markers will be automatically placed on the waveform to
indicate the measurements made.