![Agilent Technologies U2701A Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/agilent-technologies/u2701a/u2701a_user-manual_2868720089.webp)
Scope Features and Functions
2
U2701A/U2702A USB Modular Oscilloscopes User’s Guide
73
Figure 2-18
The waveform after interpolation
• Waveform persistence (infinite)
enables the oscilloscope to
update the display with new acquisitions, but does not erase
the results of previous acquisitions. Waveform persistence is
not kept beyond the display area boundary. Use infinite
persistence to measure noise and jitter, to see the worst- case
extremes of varying waveforms, to look for timing violations,
or to capture events that occur infrequently.
Clear Display
clears the previous acquisitions on the graph
display. Then, the oscilloscope will again start to accumulate
acquisitions. Turn off infinite persistence, then
press the Clear
Display key
to return the oscilloscope to the normal display
mode.
Miscellaneous
menu contains an option to display tooltips with
keyboard shortcuts. When this option is enabled, by hovering
the mouse cursor over most of the controls, a tooltip with the
appropriate keyboard shortcut to access the particular function
will appear.
U2700A_Modular_Scope.book Page 73 Tuesday, May 6, 2008 9:32 AM