Scope Features and Functions
48
U2701A/U2702A Modular Oscilloscopes User’s Guide
Trigger Options
Coupling Modes
The oscilloscope offers four types of coupling modes— Direct
Current (DC), Alternating Current (AC), Low Frequency- Reject
(LF- Reject), and High Frequency- Reject (HF- Reject).
DC coupling
allows DC and AC signals into the trigger path.
When DC coupling is selected, both DC and AC components of
the input waveform are transferred to the oscilloscope. DC
coupling enables the triggering waveforms of as low as 0 Hz
without large DC offsets to be captured.
AC coupling
enables the triggering waveforms with large DC
offsets to be captured.
LF- Reject coupling
places a 35 kHz high- pass filter in series
with the trigger waveform. The LF- Reject coupling will remove
any unwanted low frequency components from a triggered
waveform, such as power line frequencies that can interfere
with proper triggering.
HF- Reject coupling
places a low- pass filter with the 3 dB point
at 35 kHz. HF- Reject removes high- frequency noise such as AM
or FM broadcast stations from the trigger path.
U2700A_Modular_Scope.book Page 48 Tuesday, May 6, 2008 9:32 AM