![Agilent Technologies U2701A Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/agilent-technologies/u2701a/u2701a_user-manual_2868720059.webp)
Scope Features and Functions
2
U2701A/U2702A USB Modular Oscilloscopes User’s Guide
43
To configure edge triggering
1
Select the desired trigger source on the trigger source panel.
2
On the Trigger Settings panel, click on the Edge button.
3
Select the trigger level by adjusting the knob or by clicking on
or
.
4
Enter the desired holdoff time and unit, then click on Set to
proceed.
5
Select the desired slope on the Edge Trigger Settings panel.
Coupling
LF-Reject
Sets the input coupling to low frequency reject
HF-Reject
Sets the input coupling to high frequency reject
Table 2-2
Edge Trigger menu
Edge Trigger Panel Controls
N O T E
•
To trigger both edges of a clock, use
Alternating
edge mode.
•
To trigger on any activity of a selected source, use
Either
edge mode.
U2700A_Modular_Scope.book Page 43 Tuesday, May 6, 2008 9:32 AM