Maintenance
7
Testing Vapor Sensors
6
Remove the paper ball from the solvent module and dispose of it
appropriately.
7
Close the front flap again and wait a minute or so until the solvent module
alarm ceases.
8
Clear the error on the Control software
This concludes the solvent module vapour sensor test.