![Agilent Technologies PL-GPC 220 Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/agilent-technologies/pl-gpc-220/pl-gpc-220_user-manual_2867785066.webp)
66
PL-GPC 220 User Manual
6
Error Information
Errors
OVEN VAPOUR
"Spillage may
have occurred
within the oven"
Voltage < set
threshold
2 min (-, -)
Flash
Orange
YES
STBY
Immediat
e alarm
5 min (-, -)
"
YES
-
"
RI
AZ FAIL
"Attempts to
autozero the RI
have failed"
Timeout or
False Zero
20 min (10 min, 2)
Flash
Orange
NO
-
Not STBY
and not
initializing
PURGE
"The purge valve is
not functioning
correctly"
Timeout
5 s (5 s)
Flash
Red
NO
IDLE
"
INITIALIZATION
"Failed to initialize
autosampler"
cmd failed
response from
autosampler
- (-, 0)
Flash
Red
NO
-
MOVE
"Failed to move
carousel correctly"
"
- (-, 3)
Flash
Red
NO
-
VIAL LOWER
"Failed to lower
vial from the oven"
"
- (-, 0)
Flash
Red
NO
-
VIAL LIFT
"Failed to lift vial
from the carousel"
"
'- (-, 0)
Flash
Red
NO
-
SYRINGE
"Failed to move
syringe"
"
- (-, 0)
Flash
Red
NO
-
CAROUSEL LOCK
"Failed to lock
carousel after
being closed"
cmd failure
during
autosampler
run
- (-, 0)
Flash
Red
NO
-
LOAD
"Failed to switch
valve to load
position"
cmd failed
response from
autosampler
- (-, 3)
Flash
Red
NO
-
INJ
"Failed to switch
valve to inject
position"
"
- (-, 3)
Flash
Red
NO
-
Table 3
Error overview
Type
GUI Error
Message
Conditions
Response Time
(Timeout, Retries)
LED
Beeper
Default
Mode
Notes