PL-GPC 220 User Manual
4
Using the Module
Setting Up the PL-GPC 220
Changing the Injection Loop
1
Change the Idle and Low flow rates to 0.2 mL/min and the Idle temperature
to 40 °C.
2
Put the instrument into IDLE mode and let the oven cool.
Only continue if oven is at 40 °C.
3
Put the instrument into STANDBY mode and open the oven lids.
4
Replace the injection loop with the new loop size.
5
Continue with following procedure:
Preparations
The column is in RUN mode.