4
Using the Module
The Graphical User Interface
Autosampler Configuration Editor
Within the Autosampler Configuration Editor, the autosampler carousel is
graphically depicted providing an excellent visual assurance of the
programmed autosampler sequence. For clarity, the vials to be injected
(sample) are indicated in yellow and the vials to be used as rinse vials
(injection system wash) are coded in blue. The other color coding used on the
carousel are the heated zones; the red band on the circumference of the
carousel illustrates the Hot Zone and the dark blue arc the Warm Zone.
• Hot Zone
– Temperature of the samples in the Hot Zone (measured at vial
position 1). This zone spans the injection point and the area of the carousel
that is magnetically stirred should the sample vial contain a magnetic flea.
• Warm Zone
– Temperature of the samples contained in the Warm Zone
measured at vial position 19.
• Disable Heaters
– If this box is checked both Hot Zone and Warm Zone
heaters will be disabled. (Use with caution, as problems may arise when
running high temperature samples)
• Wait Until Stable
– If this box is checked, the system will wait until the oven
temperature is stable (non-flashing green LED) before performing sample
injection.
• First Vial
– Position in the carousel of the first vial in the autosampler
sequence.