PL-GPC 220 User Manual
2
Site Requirements and Specifications
Site Requirements
Site Requirements
Power Considerations
The instrument power supply has wide ranging capability. It accepts any line
voltage in the range described in
Physical Specifications
. Consequently there
is no voltage selector in the rear of the module.
Two power outlets are required; one for the main oven unit and one for the
solvent module.
For countries where 230 V will have to be installed the main power leads are
provided with NEMA-15 type plugs.
The mains supply should be fitted with an RCCB (Residual Current operated
Circuit-Breaker).
WA R N I N G
Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than specified.
➔
Connect your instrument to the specified line voltage only.
C A U T I O N
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power
line at any time.
➔
Make sure the power connector of the instrument can be easily reached and
unplugged.
➔
Provide sufficient space behind the power socket of the instrument to unplug the
cable.