PL-GPC 220 User Manual
51
Using the Module
4
Setting Up the PL-GPC 220
Changing Solvents in the PL-GPC 220
1
Change the Idle and Low flow rates to 0.2 mL/min and the Idle temperature
to 40 °C.
2
Put the instrument into IDLE mode and let the oven cool.
Only continue if oven is at 40 °C.
3
Put the instrument into STANDBY mode and open the oven lids.
4
Remove the columns from the system and replace with a barrel connector.
5
Replace the solvent reservoir with the new solvent or if an intermediary
(Acetone) solvent is required.
6
Continue with following procedure:
“Introducing Solvent to the PL-GPC
Preparations
System is in RUN mode.