80
PL-GPC 220 User Manual
7
Maintenance
Testing Overtemperature Protection Systems
This concludes the autosampler over temperature protection system test.
6
Dip the temperature sensor into the hot water.
After a few seconds the solvent module alarm buzzer will
sound and an error notification will be displayed on the
PL-GPC 220 Control software.
Next Steps:
7
Leave the probe in the hot water for a few minutes and check the temperature of the heat sink again. This should now be
noticeably cooler.
8
If the alarm condition does not occur or the heatsink does not cool down then switch off and disconnect both power
cords from the instrument.
N O T E
Contact your local Agilent Service Department for assistance.
OR
Once heat sink cooling has been verified remove the probe from the hot water and replace it in the mounting block.
9
Tighten the clamp screws slightly. Do not over-tighten.
The alarm will cease after a few seconds.
10
Clear the error on the Control software.