112
PL-GPC 220 User Manual
9
Appendix
General Safety Information
As part of the electrical safety regulations, forming part of the overall CE
compliance, the unit is fitted with a surge safe device located at the rear of the
instrument. This device protects the instrument from high voltage line surges
by filtering the incoming supply.
C A U T I O N
Solvent vapors
Liberal use of organic solvents in close proximity to the instrument may activate the
vapor sensors, causing the instrument to shutdown.
➔
Exercise care when using solvents close to the instrument.
C A U T I O N
Insulation resistance test fails
The surge safe device causes the instrument to fail the insulation resistance test of the
standard portable appliance test (PAT Test).
➔
Unplug the surge safe device before the PAT test is performed.