Introduction to PL-GPC 220
1
System Description and General Operation
Injection Sequence
1
Default condition prior to injection:
•
Injection valve in the Inject position
•
Vial in carousel
•
Syringe at mid point
2
Sample loaded
•
Valve switched to load position
•
Vial lifted to needle
•
Syringe sucks 800
μ
L from the vial to fill the loop and partially fills the
overflow loop
=di
8daY
8dajbch
Ejbe
=di
8daY
8dajbch
Ejbe