Chapter 2
29
Test Descriptions
Input VSWR
Calibrating the Vector Network Analyzer 1 from 10 MHz to 3 GHz
Step 1.
Press the
Preset
key on the Network Analyzer.
Step 2.
Set the Active Channel to 1 (press
Chan 1
).
Step 3.
Set the measurement mode to S11 (press the
Meas
key and
Reflection
).
Step 4.
Set the start and stop frequencies.
For the N8973/4/5A models (frequency range 10 MHz to 3.0 GHz): press
Start
,
1
,
0
,
M/µ
then
Stop
,
3
,
.
,
0
,
G/n
Step 5.
Set the measurement format to SWR (press the
Format
key and
SWR
).
Step 6.
Set the number of measurement points to 801 points (press the
Sweep Setup
key,
and
NUMBER of POINTS
,
8
,
0
,
1
,
x1
Step 7.
Set the power level to -30dBm (press the
Sweep Setup
key and
Power
,
-
,
3
,
0
,
x1
Step 8.
Connect the test equipment as shown in Figure 2-3.
NOTE
During this test procedure, ET model network analyzers do not require adaptor 1,
while NFA N8973A do not require adaptor 2.
Figure 2-4
Input VSWR Test Setup 1
VECTOR NETWORK
ANALYZER 1
Cable 1
Port 1
Adapter 2
(if Applicable)
Input
NOISE FIGURE
ANALYZER
Adapter 1
(if Applicable)
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...