152
Appendix D
Caring for Connectors
Mechanical Inspection: Connector Gages
Mechanical Inspection: Connector Gages
Even a perfectly clean, unused connector can cause problems if it is mechanically
out of specification. Since the critical tolerances in microwave connectors are on the
order of a few ten-thousandths of an inch, using a connector gage is essential.
Before using any connector for the first time, inspect it mechanically using a
connector gage. How often connectors should be gaged after that depends upon
usage.
In general, connectors should be gaged whenever visual inspection or electrical
performance suggests that the connector interface may be out of specification, for
example due to wear or damage. Connectors on calibration and verification devices
should also be gaged whenever they have been used by someone else or on another
system or piece of equipment.
Precision 3.5mm and SMA connectors should be gaged relatively more often than
other connectors, owing to the ease with which the center pins can be pulled out of
specification during disconnection.
Connectors should also be gaged as a matter of routine - after every 100 connections
and disconnections initially, more or less often after that as experience suggests.
Table D-1
Recommended connector gages
Connector gage kits containing all of the items required are included in many
Agilent calibration kits. They are also available separately. Part numbers are as
follows.
Type
Part Number/Ordering Information
Precision 7mm
(APC-7)
85050-80012
Precision 3.5mm
11752D
Precision 2.4mm
11752E
Type-N
85054-60047
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...