![Agilent Technologies NFA Series Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/agilent-technologies/nfa-series/nfa-series_performance-verification-and-calibration-manual_2868187035.webp)
Chapter 2
23
Test Descriptions
10 MHz Out Frequency Reference Adjustment
Step 4.
Wait for the frequency counter reading to stabilize and ensure that the adjusted value
is within the published specification. Repeat the adjustment procedure until the
correct value is obtained.
Step 5.
Press
Save
from the
Time Base
sub-menu to ensure that the values are stored in
memory.
NOTE
It is recommended that you preset the NFA and repeat the 10 MHz Frequency
Reference Accuracy Test (see page 14) to verify the adjusted DAC settings.