Chapter 2
21
Test Descriptions
10 MHz Out Frequency Reference Adjustment
Test Setup
Before starting the 10 MHz Out Frequency Reference Adjustment, connect the
equipment as shown in Figure 2-2.
Figure 2-2
10 MHz Out Frequency Reference Adjustment Test Setup
Test Procedure
NOTE
Before performing any adjustments ensure that the 10 MHz Frequency Reference
Accuracy test has been done and that the measured frequency is outside the
published specification.
NOTE
The NFA must be on and in internal frequency mode for at least 1 hour before you
start the test.
NOISE FIGURE
ANALYZER
FREQUENCY
STANDARD
UNIVERSAL
COUNTER
BNC Cable
10 MHz
Ref Out
Ref In
10 MHz
BNC Cable
Channel 1