22
Chapter 2
Test Descriptions
10 MHz Out Frequency Reference Adjustment
Step 1.
Note Counter Reading 1 for the 10 MHz Reference Accuracy Test and calculate the
actual error in Hertz.
Actual Error = 10 MHz - Counter Reading 1
Step 2.
With the actual error calculated, determine which DAC setting to adjust from the
table below.
Example 2-1
Standard NFA Actual Error
If a standard NFA Actual Error = 10,000,000 Hz - 10,000,012 Hz
the error is -12 Hz
From Table 2-4, the DAC setting must be decremented by 1 coarse setting and 5 fine
settings giving an approximate -12 Hz adjustment. Therefore, if the original settings
were, for example, Coarse = 127 and Fine = 136, then the new settings after
adjustment are Coarse = 126, Fine = 131.
NOTE
A negative error value decrements
[
⇓
]
, while a positive error value increments
[
⇑
]
.
Step 3.
Access the
Service
menus
Time Base
sub-menu and press the
System
key. Then
press the
More
,
More
and
Service
menu keys. Enter the service password
-2010
,
press
Enter
, then press
Service
and
Time Base
.
Decrement
[
⇓
]
the NFAs DAC settings by pressing
Coarse
once and
Fine
five
times.
OR
You can type in the new value on the numeric keypad and then press Enter.
Table 2-4
DAC setting adjustments
Control Setting
Standard Step Size (Hz)
Option 1D5 Step Size (Hz)
Coarse
≈
7 Hz
≈
0.1 Hz
Fine
≈
1 Hz
≈
0.0015 Hz
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...