144
Appendix D
Caring for Connectors
Introduction
Introduction
Recent advances in measurement capabilities have made connectors and connection
techniques more important than ever before. Damage to the connectors on
calibration and verification devices, test ports, cables, and other devices represent an
increasing burden in downtime and expense.
This Appendix will help you get the best performance from all coaxial microwave
connectors:
•
To know what to look for when cleaning and inspecting them, in order to
preserve their precision and extend their life.
•
To make the best possible microwave connections, improving the accuracy and
repeatability of all of your measurements, saving both time and money.
Connector Part Numbers
Refer to the latest edition of the HP RF and Microwave Test Accessories Catalog for
connector part numbers.
Handling and Storage
Microwave connectors must be handled carefully, inspected before use and when
not in use, stored in a way that gives them maximum protection. Avoid touching the
connector mating plane surfaces and avoid setting the connectors contact-end down,
especially on a hard surface.
Never store connectors with the contact end exposed. Plastic end caps are provided
with all Agilent connectors and these should be retained after unpacking and placed
over the ends of the connectors whenever they are not in use. Extend the threads of
connectors that have a retractable sleeve or sliding connector nut, then put the
plastic end cap over the end of the connector.
Above all, never store any devices loose in a box or in a desk or a bench drawer.
Careless handling of this kind is the most common cause of connector damage
during storage.
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...