![Agilent Technologies NFA Series Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/agilent-technologies/nfa-series/nfa-series_performance-verification-and-calibration-manual_2868187045.webp)
Chapter 2
33
Test Descriptions
Input VSWR
Measuring VSWR from 1500 MHz to 3000 MHz
Step 1.
Connect the test equipment as shown in Figure 2-4 on page 29.
Step 2.
Set the Network Analyzer Start/Stop frequencies from 1.5 GHz to 3.0 GHz.
Press the
Start
key then
1
,
.
,
5,
G/n
then the
Stop
key and
3, ., 0
,
G/n
Step 3.
Set the fixed frequency of the NFA to 1750 MHz (press the
Frequency/Points
key
and then
Fixed Freq
,
1, 7, 5, 0
,
MHz
).
Step 4.
Restart the measurement sweep on the DUT (press the
Restart
key).
Step 5.
Using the marker peak function on the network analyzer, determine the maximum
displayed VSWR over the measured frequency range (press the
Marker Fct
n key
and
Mkr Search [ON]
,
Search: Max
).
Step 6.
Record the maximum displayed signal frequency and VSWR on the Test Record.
Step 7.
Verify that the measured VSWR from 1500 MHz to 3000 MHz is within the
published specification.
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...