Chapter 2
27
Test Descriptions
Input VSWR
3.5mm Calibration Kit
Impedance: 50
Ω
Frequency Band: 3 GHz to
26.5 GHz
UUT: N8974A and N8975A
85052D
Cable
50 N Type (M to M)
Frequency Band: 3 GHz to
26.5 GHz
11500C
Adapter
50 N Type (F) to 7 mm
(included in 85032B N Type
Calibration Kit)
85054-60001
Adapter
2.4 mm (M) to 3.5 mm
Precision (F)
1250-1745
Adaptor
2.4 mm (M) to 3.5 mm
Precision (F)
11901C
UWave Cable
2.4 mm (M) to 2.4 mm (M),
81 cm (32 in.)
Frequency band: 3 GHz to
26.5 GHz
85133C
a. Option 004 is only required for the ET models.
b. Option 004 is only required for the ET models.
c. Part 1 of this procedure lists hard and soft keys for the 8753ET. Other network analyzers such
as the 8753D or E may have different hard and soft key labelling. Please refer to the
appropriate Users Guide for the relevant menu maps.
Table 2-6
Test Equipment required for Input VSWR test
Equipment description
Critical specification for
equipment substitution
Recommended model
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...