Chapter 2
65
Test Descriptions
Gain Measurement Uncertainty
Required Test Equipment
Test Setup
Figure 2-16
Gain Measurement Uncertainty Test Setup
Table 2-17
Equipment required for Gain Measurement Uncertainty test
Equipment description
Critical specification for equipment
substitution
Recommended model
Synthesized Sweeper
Frequency Range: 50 MHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -46 dBm
83620/30/40/50B Option
001 and 008
Adapter 1
N8973A N Type (M) to 3.5mm (F)
N8974/5A 3.5mm Precision (F) to 3.5mm (F)
1250-1744
1250-1749 or 83059B
Adapter 2
83620/30 3.5mm Precision (F) to 3.5mm (F)
83640/50 2.4mm Precision (F) to 3.5mm (F)
1250-1749 or 83059B
11901B
RF Cable
50 MHz 3.5mm (M) to 3.5mm (M), 61cm (24
in.)
11500E
Input
NOISE FIGURE
ANALYZER
SYNTHESIZED
SWEEPER
Type N Cable
RF Output
Adapter
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...