Preparing for Calibration and Performance Verification Test
Purpose of Tests
Purpose of Tests
The calibration and performance test procedures verify the electrical performance of
the Agilent Technologies N8973A, N8974A and N8975A series of Noise Figure
Analyzers (NFAs) in accordance with their published specifications.
NOTE
You do not need to access the interior of the instrument to perform the tests.
If for any reason the unit measures out-of-specification, then adjustment procedures
are supplied where applicable.
The Noise Figure Analyzers must be able to reach thermal equilibrium in
accordance with the temperature stability specifications before performing any tests
or adjustments.
NOTE
System performance is only guaranteed if each instrument in the test system/process
is within the manufacturer’s recommended calibration period.
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...