52
Chapter 2
Test Descriptions
Noise Source Supply Accuracy
Noise Source Supply Accuracy
Test Description
The Noise Source Supply Accuracy test verifies that the noise source drive supply
meets the published specifications. A Digital Multimeter is connected to the +28V
noise source supply BNC connection. The supply is then tested in the on and off
states.
Test Specification
•
Noise Off
0.0V ± 1.0V
•
Noise On
+28V ± 0.1V
Related Adjustment
None
Required Test Equipment
The following equipment is required for the Noise Source Supply Accuracy test.
Table 2-9
Equipment required for Noise Source Supply Accuracy test
Equipment description
Critical specification for equipment
substitution
Recommended model
Digital Multimeter
Input Resistance of 10 M
Ω
Accuracy of 10 mV on 100 V range
3458A
DVM BNC Adapter
BNC (F) to Dual Banana Plug (M) Adapter,
50
Ω
E9637A
BNC Lead
Length 122 cm (48 in.), Frequency 10 MHz
10503A