Operation
46
Serial BERT 12.5 Gb/s User Guide
Text
Type
Description
Default
•
ErAd
OFF
o
options:
OFF, 1E0, 1E1, 1E2, …, 1E7
Digital error injection setting, which indicates
how many bit inversions occur per second. OFF
indicates no bit errors are injected, 1E0 indicates
1 error per second, 1E7 indicates 10 million errors
per second. To find the injected BER, divide by
the clock frequency. For example, at a setting of
1E2 (100) errors per second, the BER will be 1E-8
at a clock rate of 10 GHz, or 8E-9 at 12.5 GHz. See
Table 14
.
When
Display
->Ø is selected from the display
selector, pressing the
Adjust
->Config State PLUS
button will prompt the N4962A to automatically
determine the optimum detector clock phase.
This is detailed further in Section 4.6.
•
NoData 0
o
options:
0, 1
NoData option helps with the initialization
sequence. When the source is used with the
error detector, this value should be set to 1. This
helps with a condition when all zeros are loaded
into the error detector’s input, which can
erroneously cause bad error rate numbers. This
should be set to 0 when using a different
manufacturer’s pattern generator, or a known
good signal is used with the N4962A source.
•
TD= 0010
o
Min 0010 (1 µs), max 6500 (650 µs)
The TD (“Training Delay”) option delays the start
of the error detector after the BER measurement
has been initiated. This feature is used in
conjunction with the NoData function. This delay,
Содержание N4962A
Страница 1: ...Agilent N4962A Serial BERT 12 5 Gb s User Guide...
Страница 6: ......
Страница 16: ...Getting Started 16 Serial BERT 12 5 Gb s User Guide...
Страница 24: ...N4962A System Overview 24 Serial BERT 12 5 Gb s User Guide...
Страница 36: ...System Details and Performance Specifications 36 Serial BERT 12 5 Gb s User Guide...
Страница 60: ...Operation 60 Serial BERT 12 5 Gb s User Guide...
Страница 86: ...Remote GPIB Interface 86 Serial BERT 12 5 Gb s User Guide...
Страница 88: ...Copyright Agilent Technologies 2012 Third edition May 2013 Printed in Germany...