Operation
42
Serial BERT 12.5 Gb/s User Guide
Receiver
->
On
button
& light
Light indicates error detector is ON, and:
•
Elapsed time is accumulating;
•
Total bits and errors are being
counted;
•
BER measurement is being calculated.
If detector is ON, pressing button:
•
Saves state (# bits, errors, &
seconds);
•
Turns off detector;
•
Puts detector into training mode (for
auto-synchronization of incoming bit
stream).
If detector is OFF, pressing button:
•
Resets state (# bits, errors, &
seconds);
•
Turns on detector.
OFF
Data Path
->
PRBS
¯¯¯¯
Invert
button
& light
Light indicates inverted PRBS generator
output:
•
OFF when PRBS bit stream is not
inverted;
•
ON when PRBS bit stream is inverted.
Button toggles the output inversion ON
and OFF.
Inverting the bit stream after training and
turning on the detector will result in a
BER measurement approaching 1E0.
OFF
Содержание N4962A
Страница 1: ...Agilent N4962A Serial BERT 12 5 Gb s User Guide...
Страница 6: ......
Страница 16: ...Getting Started 16 Serial BERT 12 5 Gb s User Guide...
Страница 24: ...N4962A System Overview 24 Serial BERT 12 5 Gb s User Guide...
Страница 36: ...System Details and Performance Specifications 36 Serial BERT 12 5 Gb s User Guide...
Страница 60: ...Operation 60 Serial BERT 12 5 Gb s User Guide...
Страница 86: ...Remote GPIB Interface 86 Serial BERT 12 5 Gb s User Guide...
Страница 88: ...Copyright Agilent Technologies 2012 Third edition May 2013 Printed in Germany...